DFX-8 Series Ultrasonic Flaw Detectors

115

The DFX-8+ is also equipped with an analog outputs to send a 0-10 volt electrical
signal to an external device to control a system process to respond in a different
manner considering the voltage measured. The outputs are updated at the same
rate as the PRF (pulse repetition frequency) setting. There is 100 ohm output
impedance which will vary the voltage slightly and should therefore be calibrated to
the device connected. The output options are amplitude or distance. Refer to the
Top & Bottom end caps or connections on page 24 for further info on the pins used
on the Lemo 1 5 pin connector. The procedures for activating and selecting
amplitude or distance are outlined below:

Activating Analog Out

1) Press the MENU key once to activate the menu items tab. Press the MENU

key multiple times to tab right and the ESC key multiple times to tab left until

the UTIL menu is highlighted and displaying the submenu items.

Alternatively, press the

LEFT and RIGHT arrow keys multiple times

using the left button wheel.

2) Use the

UP and DOWN arrow keys multiple times using the left button

wheel to scroll through the sub menu items until ANALOG OUTPUT is

highlighted.

3) Use the

LEFT and RIGHT

arrow keys multiple times on the right button

wheel to toggle ANALOG OUTPUT thickness/amplitude.

4) Press the MEAS key to return to the measurement screen.