DFX-8 Series Ultrasonic Flaw Detectors

109

8) Press the OK key to set the gain step and return to the menu screen, or

ESC to cancel entering the gain step.

9) Finally, press the MEAS key to return to the measurement screen.

11.17 Detect Modes

The DFX-8 has three selectable modes of detection: Z-Cross, Flank, and Peak. The
following diagram illustrates the difference between the options:

Z-Cross

Flank

Peak

Refer to the diagrams above. The current polarity is positive, as shown by the
position of the gate above the baseline:

Z-Cross: The detection occurs as the first positive half cycle crosses the baseline
and goes negative.

Flank: The detection occurs as the first half cycle left edge rises and breaks the
threshold.

Peak: The detection occurs at the highest peak in the gate.

The procedure to select the DETECT MODE is outlined below: