Dakota Ultrasonics DFX-8 plus MANUAL1 User Manual

Page 84

Advertising
background image

Dakota Ultrasonics

80

D

EFF

= Effective Diameter

K = Aspect Ratio

A = Largest Dimension

0.97 = Fixed Ratio

Calculated results can also be compared to the following table:

Ratio of the sides

B/A K

1.0 1.37

0.9 1.25

0.8 1.15

0.7 1.09

0.6 1.04

0.5 1.01

0.4 1.00

0.3 0.99

0.2 0.99

0.1 0.99

10.4 Parameter Definitions DGS(AVG)

This section outlines the parameters in the DGS/AVG toolbox, and provides a brief
definition of each parameter:

DGS: Activate the DGS on/off.

Reference Gain: Activated to set the reference gain using a given reflector from a
calibration standard.

Advertising
This manual is related to the following products: