114

Activating and Using Compare

1) Press the MENU key once to activate the menu items tab. Press the MENU

key multiple times to tab right and the ESC key multiple times to tab left until

the UTIL menu is highlighted and displaying the submenu items.

Alternatively, press the

LEFT and RIGHT arrow keys multiple times

using the left button wheel.

2) Use the

UP and DOWN arrow keys multiple times using the left button

wheel to scroll through the sub menu items until COMPARE is highlighted.

3) Use the

LEFT and RIGHT

arrow keys multiple times on the right button

wheel to toggle COMPARE on/off. Note: The COMPARE feature now

overrides the

PEAK key and its functions while activated.

4) Press the MEAS key to return to the measurement screen.

5) Acquire the waveform of the sample flaw on the display, and press the

PEAK key to freeze and send it into the background on the display in

another color. Pressing the

PEAK key a second time to will clear the

waveform.

6) Finally, repeat steps 1 – 3 to disable the COMPARE feature.