DFX-8 Series Ultrasonic Flaw Detectors

87

Note: The DGS curve uses 16 points to plot the curve.

4) Press the MENU key once to activate the menu items tab. Press the MENU

key multiple times to tab right, and the ESC key multiple times to tab left,

until the DGS menu is highlighted and displaying the submenu items.

Alternatively, press the

LEFT and RIGHT arrow keys multiple times

using the left button wheel.

5) Press the

UP and DOWN arrow keys multiple times on the left button

wheel until REFERENCE GAIN is highlighted, followed by pressing the

LEFT and RIGHT

arrow keys on the right button wheel to toggle OFF. The

DGS curve is now set and ready to be used for inspection.