DFX-8 Series Ultrasonic Flaw Detectors

71

6) Press the MENU key once to activate the menu items tab. Press the MENU

key multiple times to tab right, and the ESC key multiple times to tab left,

until the DAC menu is highlighted and displaying the submenu items.

Alternatively, press the

LEFT and RIGHT arrow keys multiple times

using the left button wheel.

7) Press the

UP and DOWN arrow keys multiple times on the left button

wheel until DRAW is highlighted, followed by pressing the

LEFT and

RIGHT

arrow keys on the right button wheel until OFF is displayed. The

DAC curves are now set and ready to be used for inspection.