Dakota Ultrasonics

64

4) Press the

LEFT and RIGHT

arrow keys multiple times on the right

button wheel to adjust the size or width of GATE1.

Note: When the signal is in the gate, the height indicator will detect and display

the % full screen height.

Setting the Level

1) Press the left button wheel, press the

LEFT and RIGHT arrow keys to

scroll through the 9 individual Hot Menus, and the

UP and DOWN arrow

keys to scroll through the Hot Menu items until the AWS BOX: cell is

highlighted. Alternatively, the MEAS & ESC keys can be used to scroll

through the Hot Menu items.

2) Press the

LEFT and RIGHT

arrow keys multiple times on the right

button wheel, to position the cursor box over the waveform.

3) Press the ENTER key to set the REFGN (reference gain).

Note: This process can be repeated and reset multiple times, or until the SET

REF GAIN feature is toggled OFF.

4) Press the MENU key once to activate the menu items tab. Press the MENU

key multiple times to tab right, and the ESC key multiple times to tab left,

until the AWS menu is highlighted and displaying the submenu items.

Alternatively, press the

LEFT and RIGHT arrow keys multiple times

using the left button wheel.

5) Press the

UP and DOWN arrow keys multiple times on the left button

wheel until SET REF GAIN is highlighted, followed by pressing the LEFT

and RIGHT arrow keys to toggle OFF.

6) Finally, press the MEAS key to return to the main measurement screen.