Dakota Ultrasonics DFX-8 plus MANUAL1 User Manual

Page 44

Advertising
background image

Dakota Ultrasonics

40

The gauge will require a few basic adjustments in order to get underway with

the calibration process. For this example, in order to avoid multiple steps for

setting the gauge features/options, the default setup will be loaded as a starting

point. However, the user should invest the necessary time getting to know the

features and options included in the DFX-8, as well as gaining a good

understanding of their function:

Loading Default Setup

1) Press the MENU key once to activate the menu items tab. Press the MENU

key multiple times to tab right and the ESC key multiple times to tab left until

the SEUTP menu is highlighted and displaying the submenu items.

Alternatively, press the

LEFT and RIGHT arrow keys multiple times

using the left button wheel.

2) Use the

UP and DOWN arrow keys multiple times on the left button

wheel to scroll through the sub menu items until DEFAULT SETUP is

highlighted.

3) Press the ENTER key to display the CONFIRM screen.

4) Press the OK key to load the DEFAULT SETUP.

Setting the Range

1) Using the left button wheel, press the

LEFT and RIGHT arrow keys to

scroll through the 9 individual Hot Menus, and the

UP and DOWN arrow

keys to scroll through the Hot Menu items, until the RANGE cell is

highlighted. Alternatively, the MEAS & ESC keys can be used to scroll

through the Hot Menu items.