DFX-8 Series Ultrasonic Flaw Detectors

49

Note: A velocity and zero value will be calculated and displayed in the fields a

shown above.

8) Finally, press the MEAS key to return to the measurement screen.

In order to trigger on anything that’s detected in the calibrated range, the

boundaries of the gate must be adjusted to include the entire range. Caution:

in both z-cross & flank detection settings, the first signal to cross the gate will be

triggered on. However, if the peak detection setting has been enabled, the

signal with the most amplitude will be detected.

Adjusting the Gate to Calibrated Range

1) Using the left button wheel, press the

LEFT and RIGHT arrow keys to

scroll through the 9 individual Hot Menus, and the

UP and DOWN arrow

keys to scroll through the Hot Menu items until the GATE1 cell is

highlighted. Alternatively, the MEAS & ESC keys can be used to scroll

through the Hot Menu items.

2) Press the

LEFT and RIGHT

arrow keys on the right button wheel to

move the gate start to a position less than 1 inch.