4) Once the desired Pulser Voltage is displayed, press the MEAS key to

return to the measurement screen.

11.5 Material Velocity Charts

The DFX-8 has two material velocity charts built into the gauge with both longitudinal
and shear material velocities. It should be noted that in order to display the chart
containing the shear wave velocities the TRIG mode must be activated, as we’ve
assumed an angle beam transducer will be used for the inspection process. The
procedure is listed below:

Selecting a Basic Material Type

1) Press the MENU key once to activate the menu items tab. Press the MENU

key multiple times to tab right and the ESC key multiple times to tab left until

the CAL menu is highlighted and displaying the submenu items.

Alternatively, press the

LEFT and RIGHT arrow keys multiple times

using the left button wheel.

2) Use the

UP and DOWN arrow keys multiple times on the left button

wheel to scroll through the sub menu items until MAT is highlighted.