DFX-8 Series Ultrasonic Flaw Detectors

83

Alternatively, press the

LEFT and RIGHT arrow keys multiple times

using the left button wheel.

2) Use the

UP and DOWN arrow keys multiple times on the left button

wheel to scroll through the sub menu items until DGS is highlighted.

3) Press the

LEFT and RIGHT

arrow keys on the right button wheel to

toggle the DGS feature ON. The menu items should now become active

and adjustable, as illustrated above.

Trigger Curve

1) Press the

UP and DOWN arrow keys multiple times on the left button

wheel until TRIGGER CURVE is highlighted, followed by pressing the

LEFT and RIGHT

arrow keys on the right button wheel to scroll the curve

options. When the correct curve is displayed, proceed to the next setting.

Effective Diameter

1) Press the

UP and DOWN arrow keys multiple times on the left button

wheel until EFF PROBE DIAMETER is highlighted, followed by pressing the

LEFT and RIGHT

arrow keys on the right button wheel to scroll the

value. When the correct diameter is displayed, proceed to the next setting.

2) Alternatively, press the ENTER key to display the Digits Edit Box.

3) Press the

UP and DOWN arrow keys multiple times on the left button

wheel to scroll the highlighted value.

4) Press the

LEFT and RIGHT arrow keys multiple times using the left

button wheel to scroll the digit locations.

5) Repeat steps 3 & 4 until the diameter is correctly displayed.

7) Press the OK key to set the EFF PROBE DIAMETER and return to the

menu screen, or ESC to cancel entering the EFF PROBE DIAMETER.