Dakota Ultrasonics DFX-8 plus MANUAL1 User Manual

Page 99

Advertising
background image

DFX-8 Series Ultrasonic Flaw Detectors

95

1) Press the MENU key once to activate the menu items tab. Press the MENU

key multiple times to tab right and the ESC key multiple times to tab left until

the CAL menu is highlighted and displaying the submenu items.

Alternatively, press the

LEFT and RIGHT arrow keys multiple times

using the left button wheel.

2) Use the

UP and DOWN arrow keys multiple times on the left button

wheel to scroll through the sub menu items until VELOCITY is highlighted.

3) Press the

LEFT and RIGHT

arrow keys on the right button wheel to

scroll the value. When the correct velocity value is being displayed,

proceed to step 8.

4) Alternatively, press the ENTER key to display the Digits Edit Box.

5) Press the

UP and DOWN arrow keys multiple times on the left button

wheel to scroll the highlighted value.

6) Press the

LEFT and RIGHT arrow keys multiple times using the left

button wheel to scroll the digit locations.

7) Repeat steps 5 & 6 until the velocity value is correctly displayed.

8) Press the OK key to set the velocity and return to the menu screen, or ESC

to cancel entering the velocity.

9) Finally, press the MEAS key to return to the measurement screen.