Note: You must set 3 points to draw a TCG curve, and a total of 16 points can

be set. Any point can be cleared by pressing the CLR key when DRAW is

active.

6) Press the MENU key once to activate the menu items tab. Press the MENU

key multiple times to tab right, and the ESC key multiple times to tab left,

until the TCG menu is highlighted and displaying the submenu items.

Alternatively, press the

LEFT and RIGHT arrow keys multiple times

using the left button wheel.

7) Press the

UP and DOWN arrow keys multiple times on the left button

wheel until DRAW is highlighted, followed by pressing the

LEFT and

RIGHT

arrow keys on the right button wheel until OFF is displayed. The

TCG curve is now set.

8) Optionally, the TCG curve can be displayed by pressing the

UP and

DOWN arrow keys multiple times on the left button wheel until CURVE is

highlighted, followed by pressing the

LEFT and RIGHT

arrow keys on

the right button wheel toggle ON.

9) Press the MEAS key to display the main measurement screen, and the TCG

curve will be displayed.