DFX-8 Series Ultrasonic Flaw Detectors

63

Setting the Reference Gain

1) Press the

UP and DOWN arrow keys multiple times on the left button

wheel until SET REF HEIGHT is highlighted, followed by pressing the

LEFT and RIGHT

arrow keys on the right button wheel to toggle the feature

ON.

2) Using the left button wheel, press the

LEFT and RIGHT arrow keys to

scroll through the 9 individual Hot Menus, and the

UP and DOWN arrow

keys to scroll through the Hot Menu items until the GATE1 cell is

highlighted. Alternatively, the MEAS & ESC keys can be used to scroll

through the Hot Menu items.

Note: A gray cursor should be visible in the waveform area of the display.

Also, since the DFX-8 has been previous calibrated and the TRIG feature has

been setup, a gate should be enabled and active.

Peak Up on the Reflector

1) Apply couplant on the calibration standard, place the transducer in the

correct area in position with the reflector, and slide the transducer around to

locate the maximum amplitude.

2) Press the

LEFT and RIGHT

arrow keys on the right button wheel to

position the start of GATE1 to trigger on the waveform.

3) Optionally, press the left button wheel, press the

LEFT and RIGHT

arrow keys to scroll through the 9 individual Hot Menus, and the

UP and

DOWN arrow keys to scroll through the Hot Menu items until the WIDTH1

cell is highlighted. Alternatively, the MEAS & ESC keys can be used to

scroll through the Hot Menu items.