The only user serviceable adjustment is the LCD contrast (see below). There
are no user serviceable components inside the instrument enclosures.

WARNING

There are potentially lethal voltages inside this
instrument’s enclosure. The source of these voltages
are from the line power and also from the System I/O
connector.

6.1 LCD Contrast Adjustment

The LCD contrast is optimized for "above the display" viewing angles and
adjusted at the factory. It may be better optimized on site for use in positions
that place the instrument in extreme viewing angles.

To adjust for best possible contrast in the installed position use a potentiometer
adjustment tool or small common screwdriver carefully inserted through the
front panel (see

section 2.4 on page 2-9

, Item 11) and turn clockwise or counter

clockwise to obtain the best possible display contrast for your viewing angle.

6.2 Error Messages

The following error codes are generated and displayed by the XTM/2.

6.2.1 Powerup Errors

ERR 0 . . . . . . Film parameters lost on power up. This may be cleared by

pressing any key. All film and layer parameters will have to be
re-entered.

ERR 2 . . . . . . Line power loss. This message is cleared by pressing any key.

ERR 9 . . . . . . Process data lost on power up. This is cleared by pressing any

key. Automatic process recovery will not be possible.

NOTE: Upon detection of power failure, all current layer and process data is

normally saved for process recovery use on subsequent deposition
system recovery.