XTM/2 Operating Manual

3.4 Sensor Selection Guide

The choice of sensor type must be dictated by the process, the deposition
material and the physical characteristics of the process chamber. General
guidelines for each sensor type produced by INFICON are outlined in the
Sensor Selection table below. For specific recommendations, consult your
INFICON representative.

*These temperatures are conservative maximum device temperatures, limited by the properties
of Teflon at higher temperatures. In usage, the water cooling allows operation in environments
that are significantly elevated, without deleterious effects.

NOTE: Do not allow water tubes to freeze. This may happen if the tubes pass

through a cryogenic shroud and the water flow is interrupted.

NOTE: For best operation, limit the maximum input water temperature to less

than 30 °C.

NOTE: In high temperature environments more heat may transfer to the water

through the water tubes than through the actual transducer. In extreme
cases it may be advantageous to use a radiation shield over the water
tubes.

Table 3-1 Sensor Selection Table

Name

IPN

Temp

°C

Crystal

Exchange

Utility

Connector

Comments

Standard

750-211-G1

130°

Front

Side

Standard
w/Shutter

750-211-G2

130°

Front

Side

Compact

750-213-G1

130°

Front

Rear

For tight spaces

Compact
W/Shutter

750-213-G2

130°

Front

Rear

For tight spaces

Dual

750-212-G2

130°

Front

Side

Two crystals for
crystal switch.
Includes Shutter

Sputtering

007-031

130°

Rear

Side

For RF and diode
sputtering. (Optional
shutter available.)

Bakeable

12" (304.8 mm)
20" (508 mm)
30" (762 mm)

007-219
007-220
007-221

450°

Front

Side

Must remove water
cooling and open the
tubes prior to bakeout

Bakeable
w/Shutter

12" (304.8 mm)
20" (508 mm)
30" (762 mm)

750-012-G1
750-012-G2
750-012-G3

450°

Front

Side

Must remove water
cooling and open the
tubes prior to bakeout

CrystalSix

750-446-G1

130°

Front

Side

6 crystals for process
security.