6.4.1 Standard and Compact

Follow the procedure below to replace the crystal in the Standard and Compact
sensor: (see

Figure 6-3

)

1

Gripping the crystal holder with your fingers, pull it straight out of the sensor
body.

2

Gently pry the crystal retainer from the holder (or use crystal snatcher; see

Figure 6-6 on page 6-19

).

3

Turn the retainer over and the crystal will drop out.

4

Install a new crystal, with the patterned electrode face up.

5

Push the retainer back into the holder and replace the holder in the sensor
body.