6 remote command – INFICON XTM/2 Thin Film Deposition Monitor User Manual
Page 78
3 - 26
IP
N 07
4-
18
6S
XTM/2 Operating Manual
3.8.5.6 Remote Command
The format of the remote command is:
R xx # . . A space is used as a delimiter between xx and #
where:
R . . . . . . Is the literal R
xx . . . . . Is the remote code per list below.
# . . . . . . Is the associated value needed for some remote commands.
R 0. . . . . Open Shutter
R 1. . . . . Close Shutter
R 2. . . . . Locks out parameters via the front panel
R 3. . . . . Unlocks parameter changes via the front panel
R 4. . . . . Zeros Thickness accumulation
R 5. . . . . Zeros Timer
R 6. . . . . Output override on. [Allows external control of relays.]
R 7. . . . . Output override off.
R 8 # . . . Set output # (if output override on). [Closes Relay #, see
R 9 # . . . Clears output # (if output override on). [See R8.]
R 10. . . . Clear power up error messages. [See S11 commands.]
R 23. . . . Set "250ms Data Ready" Service request (IEEE only).
R 24. . . . Clear "250ms Data Ready" Service request (IEEE only).