2 inputs, 3 chart recorder – INFICON XTM/2 Thin Film Deposition Monitor User Manual

Page 67

Advertising
background image

3 - 15

IP

N 07

4-

18

6S

XTM/2 Operating Manual

3.7.2 Inputs

Inputs are activated by pulling the specific input's terminal to ground (<0.8V)
through a contact closure to common (GND) or with TTL/CMOS logic having
current sink capability of 2 ma (1 low power TTL load). These ports are read
every 250 ms; signals must be present during a read cycle.

3.7.3 Chart Recorder

The chart recorder output has 12 bit resolution with one additional bit of sign
information over the range of -10 to +10 volts. It can supply up to 5 milliamps
and has an internal resistance of 100 ohms. The output is proportional to rate,
thickness or rate deviation depending on the setting of the configuration
switches; see

section 2.5.2 on page 2-15

. It is normal for ripple to appear on

these outputs to a maximum of 5 mV at ~84 Hz. This output is updated every
250 milliseconds.

Table 3-4 System I/O Connector

Pin #

Function

Description

Input #

14,15,16,17

Input Common (GND)

Used as reference for activating
any of the inputs

18

OPEN

Detection of a falling edge
duplicates front panel OPEN

1

19

CLOSE

Detection of a falling edge
duplicates front panel CLOSE

2

9

CRYSTAL FAIL
INHIBIT

Presence of a closure to ground
reference prohibits the closure
of the Crystal Fail Relay.

3

20

ZERO thickness

Detection of a falling edge
duplicates the front panel
ZERO, for thickness only.

4

21

ZERO timer

Detection of a falling edge
duplicates the front panel ZERO
for the timer only.

5

Advertising