INFICON XTM/2 Thin Film Deposition Monitor User Manual

Page 108

Advertising
background image

6 - 6

IP

N 07

4-

18

6S

XTM/2 Operating Manual

9. crystal fail always on

a. XIU/oscillator not
connected

verify proper
sensor/oscillator
connections

b. XIU/oscillator
malfunctioning

b. if available, insert a
known working
XIU/oscillator in place of the
suspect one; if
XIU/oscillator is confirmed
bad, contact INFICON
service department

c. defective cable from
feedthrough to
XIU/oscillator or from
instrument to XIU/oscillator

c. use an ohm meter or
DVM to check electrical
continuity or isolation as
appropriate

d. poor electrical contact in
the transducer,
feedthroughs, or in-vacuum
cable

d. use an ohm meter or
DVM to check electrical
continuity or isolation as
appropriate

e. failed crystal/no crystal

e. replace crystal/insert
crystal

f. two crystals placed in the
same crystal holder

f. remove one of the crystals

Table 6-1 Troubleshooting the Instrument (continued)

SYMPTOM

CAUSE

REMEDY