7 input and output details, 1 relays, Section 3.7 on – INFICON XTM/2 Thin Film Deposition Monitor User Manual
Page 66
Advertising
3 - 14
IP
N 07
4-
18
6S
XTM/2 Operating Manual
3.7 Input and Output Details
3.7.1 Relays
WARNING
The relay, relay circuit, and associated pins in the I/O
connector(s) have a maximum voltage rating of
30 V(dc) or 30 V(ac) RMS or 42 V(peak). The maximum
current rating per connector pin or relay contact is
2.5 Amps.
Their function is as follows:
Table 3-3 System I/O Connector
Pin #
Function*
Closed Contacts
Open Contacts
Relay #
1,2
Source Shutter
During "Shutter Open"
State.
Balance
1
3,4
Thickness
Setpoint
When SPT
THICKNESS is
exceeded.
Balance
2
5,6
Timer Setpoint
When SPT TIMER is
exceeded.
Balance
3
7,8
Crystal Fail
When crystal fails to
oscillate.
Balance
4
Function may be overwritten by Remote Communications Commands "R6 - R9", see
.