Generally, install the sensor as far as possible from the evaporation source (a
minimum of 10" or 254 mm) while still being in a position to accumulate
thickness at a rate proportional to accumulation on the substrate.
shows proper and improper methods of installing sensors.
To guard against spattering, use a source shutter or crystal shutter to shield the
sensor during the initial soak periods. If the crystal is hit with even a minute
particle of molten material, it may be damaged and stop oscillating. Even in
cases when it does not completely stop oscillating, it may become unstable.
Figure 3-4 Sensor Installation Guidelines