INFICON XTM/2 Thin Film Deposition Monitor User Manual

Page 77

Advertising
background image

3 - 25

IP

N 07

4-

18

6S

XTM/2 Operating Manual

S 10 . . . . Present configuration switch settings—returns 16 ASCII bytes with a

value of 0 or 1, corresponding to the position of switches 1-16. Byte
1 corresponds to switch 1. See S13 also.

S 11 . . . . Power-up errors

S11 Response Codes

0. . . . Parameter data checksum error—indicates a loss of stored

parameter data.

1. . . . STBY/ON sequence since last query—the front panel power

switch has been used since the last inquiry.

2. . . . Line power failure.

9. . . . Process data checksum error—indicates a loss of process

data.

10. . . No errors.

NOTE: If more than 1 error condition exists, the response string will

list them all, each separated by a single space.

NOTE: STBY/ON status is cleared automatically by issuing an S11

command. All others require intentional clearing (available
via remote command).

S 12 . . . . Datalog output, see

section 3.8.4 on page 3-21

. The data is

separated by a space instead of CR LF.

S 13 . . . . Instrument Configuration, the position of the configuration switches

at the last STBY/ON sequence. Use this command to determine the
instrument’s current operating configuration. See S10 also.

S10 Response Strings

Byte 16 15 14 13 12 11 10 9

8

7

6

5

4

3

2

1

[1 = switch on, refer to

section 2.5.2 on page 2-15

]