2 state descriptions – INFICON XTM/2 Thin Film Deposition Monitor User Manual

Page 87

Advertising
background image

4 - 3

IP

N 07

4-

18

6S

XTM/2 Operating Manual

4.2 State Descriptions

Operating the XTM/2 as a film thickness/rate monitor requires programing the
film parameters, refer to

section 2.6 on page 2-22

. A film sequence begins with

a OPEN command and ends when the film in process reaches the FINAL
THICKNESS
or the CLOSE switch is pressed. The RATE and THICKNESS
displayed is modified by the values programmed in the possible parameters
and the units used are set by the configuration switches, refer to

section 2.5.2

on page 2-15

.

In reviewing the state diagram of

Figure 4-2 on page 4-2

, note that there are

only two basic machine states, SHUTTER OPEN or SHUTTER CLOSED. On
power-up, the machine proceeds to the SHUTTER CLOSED state after
performing a series of memory checks. Various abnormal conditions are
indicated by ERR# messages, see

section 6.2.1 on page 6-1

and

section 6.2.2

on page 6-2

. The operations indicated by a pointing finger are manual key

strokes or the equivalent system hardware input or computer communications
command.

In addition, there are two basic display modes: Operate and Program. Pressing
the PROG switch at any time alternates between these modes.

Information may be stored and recalled for nine setups. The particular setup is
chosen by moving the cursor to the FILM parameter and selecting 1-9. In
addition, Film 0 is preprogrammed for Test Mode use, refer to

section 3.6 on

page 3-11

.

Table 4-1 State Descriptions

State

Condition

Source
Shutter

1. SHUTTER CLOSED

Will accept an OPEN command.

Open

2. SHUTTER OPEN

Will accept a CLOSE command.

Closed