6.

Chart Recorder; Model 1327.
The Model 1327 chart recorder may be positioned on the rack shelf or on the bench.

a.

Connect System Cable 2 between the BioLogic LP Controller and the recorder as follows:

• The mini-DIN connector is connected to the port marked “UV Chart” on the rear of the

BioLogic LP Controller.

• The DIN connector is connected to the single DIN port on the side of the chart recorder.

This cable provides system control of pen up/down, event marks and paper advance, but chart
speed MUST be set on the recorder faceplate itself.

b.

Conductivity signals are recorded on channel 2 of the recorder using System Cable 4 as follows:

• The mini-DIN connector is connected to the port marked “Cond. Chart” on the rear of the

BioLogic LP Controller.

• The banana plugs are connected to the ports marked CH 2 on the side of the recorder (red

wire to +, black wire to ground | ).

c.

Set both channel inputs on the recorder to 1V. Set all other switches to their position marked in
green. Connect the power adapter to a wall outlet, and plug the power lead into the chart recorder.

7.

Fraction Collector; Model 2110.

a.

The Model 2110 fraction collector may be positioned on the rack shelf or on the bench.

b.

Connect System Cable 1 between the BioLogic LP Controller’s Fraction Collector port and the frac-
tion collector as shown in Figure 11 and described below:

• The 9-pin D connector is connected into the single D port on the rear of the Model 2110.

• Connect the power cable. (A power cable is available to connect the fraction collector to the

BioLogic LP Controller, eliminating the need for a wall outlet.)

c.

Position the SV-3 valve on the rack using the integral clamp.

d.

Connect the SV-3 valve cable mini-DIN connector to the Diverter Valve port.

8.

Fraction Collector; Model 2128.

a.

Place the Model 2128 fraction collector on the bench, next to the BioLogic LP.

b.

Connect System Cable 15 between the BioLogic LP Controller’s Fraction Collector port and the
fraction collector as shown in Figure 11. The 15-pin D connector is connected into the single D port
on the rear of the Model 2128.

c.

Connect System Cable 3 between the BioLogic LP Controller’s Diverter Valve port and the fraction
collector as shown in Figure 11. The mini-DIN connector is connected into the single mini-DIN port
on the rear of the Model 2128.

d

Connect the power cable. (A power cable is available to connect the fraction collector to the
BioLogic LP Controller, eliminating the need for a wall outlet.)

When all the connections have been made, turn the units around. Plug all the power cables into an
approved power-strip. The system can now be plumbed.