Reference description – Bio-Rad BioLogic LP System User Manual

Page 52

Advertising
background image

Table 10. (continued)

Manual Mode Operation: Conductivity Monitor

Reference

Description

Min/Max: Allows you to set a range, from 0.00 mS to 999.99
mS, not selectable under Set Range’s incremental values. To
use this function, run buffer A through the system until the con-
ductivity reading stabilizes, and note the conductivity reading.
Then run buffer B through the system and note its conductivity
reading. In the conductivity Min/Max screen, enter the noted A
and B values as the minimum and maximum.

45

System Operation

Method: <<name>>

Manual Operation

Conductivity Monitor: 0.0111 mS

Range: Min: 0.00 mS

Max: 5.00 mS

(set min)

SET MAX

CANCEL

OK

M3

MU

MU

(Set the Minimum and

Maximum values for

the range of detection)

Method: <<name>>

Manual Operation

Conductivity Monitor:

0.00 mS

Range:

500.0 mSFS

SET RANGE

CALIBRATE

MIN/MAX

Conductivity Cell Constant:

40.0000

(constant)

SOLUTION

CANCEL

OK

Enter the Conductivity of the solution
now in the Conductivity Flow Cell:

0.00 mS/cm

CONSTANT

(solution)

CANCEL

OK

(return to MC)

(return to previous screen)

(return to MC)

(return to MC)

Method: <<name>>

Manual Operation

Conductivity Monitor:

0.0111 mS

Range: 500.0 mSFS

INCREASE

DECREASE

CANCEL

OK

Method: <<name>>

Manual Operation

Conductivity Monitor:

0.0111 mS

Range: Min: 0.00 mS

Max: 5.00 mS

(set min)

SET MAX

CANCEL

OK

press to increment/

decrement the range value

by approximately 50% each time

the button is pressed

(return to MC)

(Set the

maximum value)

(set minimum

value is

selected)

MC

M2a

MC

MANUAL MODE: CONDUCTIVITY

SET CONDUCTIVITY RANGE

CALIBRATE FOR CONDUCTIVITY

M1

M2

SET MIN/MAX VALUES FOR CONDUCTIVITY

M3

M1

M3

M2

LEGEND

Figure 19. Manual Operation of the Conductivity Monitor

(Screen displays are keyed to the discussion in Table 10, Manual Mode Operation : Conductivity Monitor)