Bio-Rad BioLogic LP System User Manual

Page 89

Advertising
background image

5.4.2.3 Pausing a Run

Runs can only be Paused by pressing the Pause softkey on the Run Screen they cannot be programmed to
occur.

Table 26.

Run Mode: Run Pause

Reference

Description

During a Pause, all instruments stop functioning and the
elapsed time or volume will not change. The Resume Run
softkey concludes the Pause by starting all instruments and the
elapsed time or volume counter.

During a Pause, the method itself can be changed before the
run is resumed. This is accomplished by pressing the Program
mode key to enter Program mode. The Run mode key light
starts flashing as a reminder that a run in progress has been
paused. While in program mode, the method can be edited
normally with a single important exception. The exception is
that the parts of the method which have already occurred can-
not be changed. The pump step in progress can be edited, but
doing so will split the step into two steps: the part of the step
which has already occurred will be the first of the two steps,
and the part of the step which has not yet occurred will be the
second step. As a result of the split, the number of steps in the
method will increase by one. After editing the method, a choice
will be presented to either replace the original method, to save
the modified method using a new name, or to finish the run with
the modified method not saved. Completing this choice returns
to the Run Paused screen.

For information describing how to edit the method, and the
naming and saving options, see section 5.3 Programming
Mode.

82

System Operation

Method: <<name>>

Volume: 0.51 ml

Pump: 1.00 ml/min A

UV:

-0.0208 AU

Frac: Collecting

Cond:

0.10 mS

RESUME RUN

STOP RUN

(confirm and exit

Run mode)

R2

R2c