0 system plumbing, 1 general guidelines for plumbing the system – Bio-Rad BioLogic LP System User Manual

Page 29

Advertising
background image

4.0 SYSTEM PLUMBING

This chapter begins with discussion of plumbing practices and general guidelines, followed by a general pro-
cedure for plumbing the major system components. Later sections discuss how to plumb Bio-Rad’s low
pressure valves.

4.1 GENERAL GUIDELINES FOR PLUMBING THE SYSTEM

The BioLogic LP uses the following tubing:

System tubing: 0.8mm ID or 1.6mm ID Tygon tubing.

Pump Tubing: 0.8mm ID, 1.6mm ID, or 3.2mm ID PharMed tubing. Refer to the following table
when selecting a tubing.

Warning
Do not use pump tubing with wall thickness greater than 1.0 mm. Using tubing with a wall thickness
greater than 1.0 mm can damage the pump and void the warranty.

Table 4

Comparison of Flow Rate Ranges for Different Tubing IDs

22

System Plumbing

1.0

.02

0.8/0.8 mm

0.8 mm: 0.05 to 1.75 ml/min
1.6 mm: 0.1 to 6.5 ml/min
3.2 mm: 0.2 to 20.0 ml/min

FLOW RATE RANGE ml/hr

FLOW RATE RANGE ml/min

OPTIMAL RANGE

MAXIMUM RANGE:

TUBING SIZE

(ID/WALL)

1.6/0.8 mm

3.2/0.8 mm

.04

.06 .08

.2

.4

.6 .8

2

4

6 8

20

3.0

5.0

10

30

50

100

300 500

1000