Table 22. (continued) program mode’s alarm table – Bio-Rad BioLogic LP System User Manual
Page 81

Table 22. (continued)
Program Mode’s Alarm Table
Reference
Description
Alarm Settings: Enter the time or volume for the alarm to occur
using the number keys.
•
Hold On: This programs the method to Hold when the
alarm occurs.
•
Hold Off: This programs the method run normally when the
alarm occurs.
•
Cancel: This returns to the Alarm Table without added or
modifying the alarm.
•
OK: This returns to the Alarm Table with the new or modi-
fied alarm in the table.
74
System Operation
A1a
Method: Volume: 1.00 ml
Alarm: 1
Volume:
0.00 ml
Hold Method: NO
HOLD ON
(hold off)
CANCEL
OK
Changes Hold
from NO to YES
Method will
not be
stopped by
alarm
(returns to
A1)
(returns to
A1)
Figure 30. Programming: the Alarm Table
(Screen displays are keyed to the discussion in Table 22)
>Alarm 1 Not Programmed
Alarm 2 Not Programmed
Alarm 3 Not Programmed
ADD
OK
Method:
Volume: 1.00 ml
Alarm: 1
Volume: 0.00 ml
Hold Method: NO
HOLD ON
(hold off)
CANCEL
OK
>Alarm 1 1.00 ml Hold Method: Yes
Alarm 2 Not Programmed
Alarm 3 Not Programmed
EDIT
DELETE
OK
A1
A1a
A1
ALARM TABLE
A1a
PROGRAM ALARM
LEGEND
return to
Method List
return
to A1