2 manual mode operation of the fraction collector – Bio-Rad BioLogic LP System User Manual

Page 45

Advertising
background image

38

System Operation

Table 6.

Manual Mode Operation: Pump

Reference

Description

Buffer screen: Selecting Buffer from the Main Menu screen
displays the Buffer screen, shown to the left. This screen
allows you to select the buffer to be delivered to the column.
Use the Prev and Next keys to select a single buffer (A through
E) or a mixture of buffers A and B. To set a mixture, select the
Mix soft key and enter the percent of buffer B. Note: If a valve
is not connected on the BioLogic LP, its letter or letters appear
in lower case and cannot be selected.

5.2.2 Manual Mode Operation of the Fraction Collector

Table 7 discusses each of the functions available in Manual mode operation of the fraction collector. An
overview of the Fraction Collector’s Manual mode operation is shown in Figure 16.

Table 7.

Manual Mode Operation: Fraction Collector

Reference

Description

To begin Manual mode operation of the fraction collector:
a.

From the Mode keys select Manual.

b.

From the Instrument keys select Collector.

c.

The main menu for manual mode operation of the fraction
collector is shown below.

The display shows the following:

Fraction Size: The fraction size can be specified as a vol-
ume (ml) or time (minutes).

UV: The current UV reading.

Diverting to Waste/ Collect: Indicates the position of the
SV-3 Diverter valve.

Cond: The current Conductivity reading.

Description of the soft key choices follows.

Method: <<name>>

Manual Operation

Buffers: M [A] B C D E
(Select MIX for Mixture of A and B)

MIX

CANCEL

OK

M4

M4a

M3

M1

Method: <<name>>

Manual Operation

Fx Size: 1.00 ml

UV:

0.0111 AU

Diverting to Waste

Cond:

0.000 mS

START

F/X SIZE

ADVANCE

MODEL

MF

M3

M1

M2

M4

Manual

Collector