System Operation

5.2.6 Manual Mode Operation of the Valves

Table 11 discusses each of the functions available in Manual mode operation of the Valves. Figure 20 pro-
vides an overview of those functions.

Table 11.

Manual Mode Operation: Valves

Reference

Description

To manually control the Valves:
a.

From the Mode keys select Manual.

b.

From the Instrument keys select Valves.

c.

The main menu for manual control of the Valves is shown
below.

The display shows the following:

Buffer: The currently selected buffer (A - E) or mixture of
Buffers A and B.

Divert Valve: The status of the SV-3 Diverter valve.

Bypass Valve: The status of the SV-3 Bypass valve.

Description of the soft key choices follows.

Note: If the Divert or Bypass valves are not connected, their soft-
keys are displayed in lowercase letter and cannot be selected.

Buffer screen: Allows you to select the buffer delivered to the
column. You can select buffers A through E, or a mixture of
buffers A and B. Use the Prev or Next keys to select a buffer;
to select a mixture of buffers A and B, press the MIX soft key.
(Refer to the Mix screen shown below.)

Note: If a valve is not connected to the controller it’s letter or
letters appear in lower case and cannot be selected.

Mix screen: Enter the percentage of Buffer B such that
%A + %B = 100%.

Method: <<name>>

Manual Operation

Buffer: A

Divert Valve: Collecting

Bypass Valve: Column In Line

BUFFER

DIVERT

BYPASS

MV

M3

M1

M2

Method: <<name>>

Manual Operation

Buffers: M [A] B C D E

(Select MIX for Mixture of A and B)

MIX

CANCEL

OK

M1

MV

M1a

MV

Method: <<name>>

Manual Operation

Enter Mixture: 50% B

(Mixture will contain buffers A and B)

CANCEL

OK

M1a

M1

MV