Bio-Rad BioLogic LP System User Manual

Page 96

Advertising
background image

3.

Remove the screws holding the optics module together and remove the bottom half of the case.

4.

Pull the UV lamp out of its holder and unplug it from its connector.

When inserting the new UV lamp, never handle the quartz surface of the lamp; grease and fingerprints will
damage the lamp. This procedure is also discussed in the instruction sheet for the replacement lamp.

Note: When re-assembling the case, be sure the O-ring is properly seated. An incorrectly seated O-ring will
cause the unit to leak light and give poor performance.

Figure 32. UV Optics Module

89

Maintenance and Troubleshooting

O-RING

RETAINING
CLIP

LAMP

CONNECTOR

UV LAMP

FILTER

TRAY

FLOW

CELL

LAMP

HOLDER

RACK
MOUNT

RACK

MOUNT