Calibrating the temperature factor – Checkline TI-MINIMAX User Manual

Page 100

Advertising
background image

Dakota Ultrasonics

94

Determination of the Temperature Factor is accomplish by solving the equation:

Where:

6

0

0

10

)

(

)

(

×

=

T

T

L

L

L

T

T

T

p

Temperature Factor

L

0

= Ultrasonic Length at Beginning Temperature (T

0

)

L

T

= Ultrasonic Length at Ending Temperature (T)

T

0

= Beginning Temperature

T = Ending Temperature

Calibrating the Temperature Factor

1) Stabilize the sample bolt at approximately 48

0

F. If the water bath

method is used this is accomplished in a stirred bath of ice and water.

Measure and record this minimum temperature as T

0.

Note: The following steps assume that the user has performed the steps in

the previous section and calibrated the velocity. Therefore, the velocity has

been adjusted, approximate length already entered, and the MINI-MAX is

currently setup and ready to measure in elongation mode.

2) Apply a drop of couplant to the bolt or transducer, and attach it to one

end of the bolt. Rotate the transducer clockwise and counter clockwise

applying a small amount of pressure to eliminate any excess couplant

between the transducer and bolt. Be sure to always place the

transducer in the same location. This will help to eliminate any potential

measurement errors caused by changing the sound path.