MINI-MAX Ultrasonic Bolt Tension Monitor

3) Record the ultrasonic length “L

0

” of the sample bolt at the above

minimum temperature - 48

0

F.

4)

Repeat step 1 - 3 at the target temperatures (T

1

through T

5

) – 48, 68,

88, and 108

0

F (other similar range with 5 temperatures), measuring the

ultrasonic lengths (L

1

through L

5

) respectively. This process must be

done for each of the sample bolts in the experiment.

5) Use the equation below to calculate a linear regression, using the

temperatures and lengths recorded for all the sample bolts to determine

the temperature factor.

( )( )

( )

6

2

2

10

=

n

x

x

n

y

x

xy

eFactor

Temperatur

x = temperature, y = change in reference length, n = number of sample points per bolt