Setting up the mini-max for measurement – Checkline TI-MINIMAX User Manual

Page 78

Advertising
background image

Dakota Ultrasonics

72

24) Press the

UP and DOWN arrow keys to scroll through the Two Point

Zero list until TEMP COEF is highlighted.

Note: This refers to the temperature coefficient of the probe zero cal bars.

25) Press the

ENTER key to display the Digits Edit Box.

26) Press the

UP and DOWN arrow keys to scroll the highlighted value.

27) Press the

LEFT and RIGHT arrow keys to scroll the digit locations.

28) Repeat steps

20 & 21 until the TEMP COEF value is correctly

displayed.

29) Press the

OK key to return to the Two Point Zero list, or ESC to cancel

entering the

TEMP COEF.

30) Press the

ESC key once to leave the Two Point Zero list.

31) Proceed to the next section.

Before the probe zero is calculated the zero points of the cal bars need to be
ultrasonically measured and documented. This procedure is demonstrated in the
next section.

Setting up the MINI-MAX for Measurement

1) With the menu items still active from the previous section, press the

MENU key multiple times to tab right and the ESC key multiple times to

tab left until the

GEOM menu is highlighted and displaying the submenu

items.