MINI-MAX Ultrasonic Bolt Tension Monitor

8) If both parameters will be edited,

repeat steps 4 – 7.

9) Use the

UP and DOWN arrow keys to scroll to and highlight SAVE

SETUP.

10) Press the

ENTER key to activate the Setup List Box.

11) Use the

UP and DOWN arrow keys to scroll through the setups until the

target location to save the

Setup is highlighted.

12) Press the

OK key to activate the confirmation screen.

13) Press the

OK key to save the Setup, or ESC to cancel saving the

Setup.

14) Finally, press the

MEAS key to return to the measurement screen.

Note: The Name and Comments of the Setup can be edited at any time

by simply repeating the

Save Setup routine described above. Therefore,

the Save Setup function can also be considered an Edit Function.

15.4 Using the Default Setup

The default setup feature was added to the MINI-MAX to use, as a last resort, if
there are no setups stored in the gauge –factory or otherwise. The only time this
might possibly occur is if the setup file in the MINI-MAX was somehow corrupted,
and the user does not have access to a computer to re-load the factory setup