Checkline TI-MINIMAX User Manual

Page 80

Advertising
background image

Dakota Ultrasonics

74

11) Use the

UP and DOWN arrow keys to scroll through the sub menu

items until

AUTO SET is highlighted.

Note: Alternatively, the Auto Set key on the keypad can be pressed.

12) Apply a drop of couplant to the bolt or transducer, and attach it to one

end of the bolt. Rotate the transducer clockwise and counter clockwise

applying a small amount of pressure to eliminate any excess couplant

between the transducer and bolt surface. Be sure to always place the

transducer in the same location. This will help to eliminate any potential

measurement errors caused by changing the sound path.

13) Press the

ENTER key and the MINI-MAX will locate and measure the

Ultrasonic Length of the first cal bar.

14) Press the

MENU key once to activate the menu items tab. Press the

MENU key multiple times to tab right, and the ESC key multiple times to

tab left, until the

CAL menu is highlighted and displaying the submenu

items.

15) Use the

UP and DOWN arrow keys to scroll through the sub menu

items until

MEASURE ZERO is highlighted.

16) Press the

ENTER key to display the Two Point Zero list box.

17) Press the

UP and DOWN arrow keys to scroll through the Two Point

Zero list box until ULTRASONIC 1 is highlighted.

Note: ULTRASONIC 1 is the ultrasonic length of the first calibration

standard.

18) Press the

ENTER key to insert the ultrasonic length being measured in

the

ULTRASONIC 1 field.