MINI-MAX Ultrasonic Bolt Tension Monitor

9) Press the

LEFT and RIGHT arrow keys to scroll the digit locations.

10) Repeat steps

8 & 9 until the PHY LEN 1 value is correctly displayed.

11) Press the

OK key to return to the Two Point Zero list, or ESC to cancel

entering the

PHY LEN 1.

12) Press the

UP and DOWN arrow keys to scroll through the Two Point

Zero list until PHY LEN 2 is highlighted.

Note: PHY LEN 2 is the actual physical length of the second calibration

standard.

13) Press the

ENTER key to display the Digits Edit Box.

14) Press the

UP and DOWN arrow keys to scroll the highlighted value.

15) Press the

LEFT and RIGHT arrow keys to scroll the digit locations.

16) Repeat steps

14 & 15 until the PHY LEN 2 value is correctly displayed.

17) Press the

OK key to return to the Two Point Zero list, or ESC to cancel

entering the

PHY LEN 2.

18) Press the

UP and DOWN arrow keys to scroll through the Two Point

Zero list until VELOCITY is highlighted.

Note: This refers the velocity of the probe zero cal bars.

19) Press the

ENTER key to display the Digits Edit Box.

20) Press the

UP and DOWN arrow keys to scroll the highlighted value.

21) Press the

LEFT and RIGHT arrow keys to scroll the digit locations.

22) Repeat steps

20 & 21 until the VELOCITY value is correctly displayed.

23) Press the

OK key to return to the Two Point Zero list, or ESC to cancel

entering the

VELOCITY.