Checkline TI-MINIMAX User Manual

Page 65

Advertising
background image

MINI-MAX Ultrasonic Bolt Tension Monitor

should not be changed, but the temperature value should be. Therefore the bolt
becomes sort of temperature sensor in itself. This is also a good way to monitor
temperature changes in the bolt. The procedures below demonstrate how to
adjust the temperature to match the zero value.

Adjusting the Temperature to Match the Zero Value

Note: User has selected this section because the MINI-MAX, transducer,

or transducer cable has not changed and the zero value does not match.

Note: Let’s assume that the group containing the probe zero calibration

data is not the current group open. Therefore, the procedures that follow

will demonstrate how to open the probe zero calibration group, verify the

zero match, and make adjustments accordingly.

1) Press the

MENU key once to activate the menu items tab. Press the

MENU key multiple times to tab right and the ESC key multiple times to

tab left until the

DATA menu is highlighted and displaying the submenu

items.

2) Use the

UP and DOWN arrow keys to scroll through the sub menu items

until

OPEN is highlighted.

3) Press the

ENTER key to display the Group List Box.