MINI-MAX Ultrasonic Bolt Tension Monitor

Now that a bolt group has been created and material type selected to save and
document the probe zero data, this next section will go through the process of
obtaining the information needed to use the standard bolt for a probe zero
calibration bar.

Setting up a Bolt Standard – For the First Time

1) With the menu items already active, press the

MENU key multiple times

to tab right and the

ESC key multiple times to tab left until the CAL

menu is highlighted and displaying the submenu items.

2) Use the

UP and DOWN arrow keys to scroll through the sub menu items

until

ZERO MODE is highlighted.

3) Use the

LEFT and RIGHT arrow keys to scroll the zero mode options

until the

FIXED option is displayed.

Note: The fixed option sets the delay factor to zero. In other words,

disables the zero feature altogether.

4) Press the

MENU key multiple times to tab right and the ESC key

multiple times to tab left until the

MATL menu is highlighted and

displaying the submenu items.