Saving the group – Checkline TI-MINIMAX User Manual

Page 75

Advertising
background image

MINI-MAX Ultrasonic Bolt Tension Monitor

Saving the Group

Once all the parameters are set, the user has the option of saving or

canceling the new group.

5) Use the

UP and DOWN arrow keys to scroll through the new Group

List Items until CREATE NEW GROUP? is highlighted.

6) Press the

ENTER key to accept the group parameters, and activate the

confirmation screen.

7) Press the

OK key to save the New Group, or the ESC key to cancel the

New Group setup and return to the DATA menu.

Now that the bolt group has been created to document the probe zero data,
we’re ready to get started with the probe zero process. The first thing to note is,
the procedures in the next section demonstrate a

Two-Point calibration.

However, the same procedures apply to the

One-Point calibration option also.

The user would simply follow the same procedures and ignore references to the
second point.

Measuring Calibration Standards – For the First Time