Checkline TI-MINIMAX User Manual

Page 76

Advertising
background image

Dakota Ultrasonics

70

1) With the menu items still active from the previous section, press the

MENU key multiple times to tab right and the ESC key multiple times to

tab left until the

CAL menu is highlighted and displaying the submenu

items.

2) Use the

UP and DOWN arrow keys to scroll through the sub menu items

until

ZERO MODE is highlighted.

3) Use the

LEFT and RIGHT arrow keys to scroll the zero mode options

until the

TWO POINT option is displa yed.

Note: The two-point option allows the user to use two separate points for

the probe zero calibration. Note: The one point option can optionally be

selected if only one cal bar will be used. Although the procedures below

are described with the two-point option in mind, they will also work in the

case of a one-point calibration.

4) Use the

UP and DOWN arrow keys to scroll through the sub menu items

until

MEASURE ZERO is highlighted.

5) Press the

ENTER key to display the Two Point Zero list box.

6) Press the

UP and DOWN arrow keys to scroll through the Two Point

Zero list until PHY LEN 1 is highlighted.

Note: PHY LEN 1 is the actual physical length of the first calibration

standard.

7) Press the

ENTER key to display the Digits Edit Box.

8) Press the

UP and DOWN arrow keys to scroll the highlighted value.