Checkline TI-MINIMAX User Manual

Page 94

Advertising
background image

Dakota Ultrasonics

88

transducer in the same location. This will help to eliminate any potential

measurement errors caused by changing the sound path.

16) Press the

ENTER key and the Ultrasonic Length of the bolt will be

displayed.

Note: Alternatively, the Auto Set key can be pressed to activate the Auto

Set routine.

17) Calculate the correct velocity using the following equation:

ocity

CorrectVel

ocity

CurrentVel

Length

Ultrasonic

ngth

PhysicalLe

=

*

18) Press the

MENU key multiple times to tab right, and the ESC key

multiple times to tab left, until the

MATL menu is highlighted and

displaying the submenu items.

19) Use the

UP and DOWN arrow keys to scroll through the sub menu

items until

VELOCITY is highlighted.

20) Press the

ENTER key to display the Digits Edit Box.

Note: Enter the Correct Velocity calculated above.

21) Press the

UP and DOWN arrow keys to scroll the highlighted value.

22) Press the

LEFT and RIGHT arrow keys to scroll the digit locations.

23) Repeat steps

21 & 22 until the VELOCITY value is correctly displayed.

24) Press the

OK key to set the Velocity and return to the menu screen, or

ESC to cancel entering the Velocity.