MINI-MAX Ultrasonic Bolt Tension Monitor

1) Press the

MEAS key once to activate measure menu items. Press the

MEAS key multiple times to move right and the ESC key multiple times

to move left until the

LOC cell is highlighted.

2) Press the

ENTER key to display the Group View Box.

3) Press the

UP, DOWN, LEFT, and RIGHT arrow keys to scroll the target

cell cursor to the desired storage location.

Note: The first column must be used to store reference lengths.

4) Press the

ENTER key to save the current reading in the highlighted cell

location.

5) Abort the

Grid View Box by pressing the MEAS key at any time.

6) Proceed to the next section to perform and store the first

Probe Zero

verification. DO NOT REMOVE THE TRANSDUCER FROM THE

BOLT.

The next two sections outline the procedures of how to perform a probe zero
calibration by calculating a zero value, or alternatively adjusting the temperature
value to match the probe zero value. When should I use the

Calc Zero feature

versus adjusting the

Temperature, and why?

The only time the

Calc Zero feature should be used is when a different MINI-

MAX, transducer, or transducer cable is being used, other than the one initially
used to obtain the probe zero data for the probe zero calibration bolt. In this
case, calculating the probe zero value forces the new hardware to match with the