Note: Since this is the 1

st

verification following the initial measurement of

the

Probe Zero Cal Bars, the cursor should be located on a cal standard in

column B.

20) If the elongation is +/- .0001” and “In Spec”, Press the

ENTER key to

save the current reading in the highlighted cell location, storing all the

current probe zero verification data.

Repeat steps 6 - 31 for the other

probe zero calibration standards. If the probe zero is “Out of

Spec”, proceed to the next step.

21)

Assuming the elongation is “Out of Spec”, Press the MEAS key once to

activate the measure menu items. Press the

MEAS key multiple times

to move right and the

ESC key multiple times to move left, until the

TEMP cell is highlighted.

22) Use the

LEFT and RIGHT arrow keys to adjust the temperature value

util the elongation is 0.0000”, or “In Spec” at +/- 0.0001”.

23) . Press the

MEAS key multiple times to move right and the ESC key

multiple times to move left, until the

LOC cell is highlighted.

24) Press the

ENTER key to display the Group View Box. Notice the

elongation value is now “In Spec”.

25) Press the

ENTER key to save the current reading in the highlighted cell

location, storing all current probe zero verification data.

26)

Repeat steps 6 - 31 for the other probe zero calibration standards.

Once again, If the probe zero is “Out of Spec”, repeat steps 32 – 36.