3 performing a field calibration – Checkline TI-MINIMAX User Manual

Page 106

Advertising
background image

Dakota Ultrasonics

100

Applied

Load

(Measured

by Tensile

Machine)

Elongation (in time as measured by MiniMax)

At low loads, vector
can be more accurate

Vector

Regression

11.3 Performing a Field Calibration

This section outlines the necessary procedures to perform a field calibration to
determine a Load Factor and Offset (in the case of using regression). In the
calibration procedures below, three sample bolts will be used with three loads
applied to each bolt (1/3, 2/3, & Max). This should be considered the minimum
requirement when performing a field calibration. The following procedures also
assume that the user has performed the probe zero operation and the MINI-MAX
is setup and ready for measurement.

Creating a Group Name to Document Field Calibration Data