3 semi automatic mode – Checkline TI-MINIMAX User Manual

Page 88

Advertising
background image

Dakota Ultrasonics

82

2) Use the

UP and DOWN arrow keys to scroll through the sub menu items

until

TEMP MODE is highlighted.

3) Use the

LEFT and RIGHT arrow keys to scroll the temperature mode

options. Once the

MANUAL temperature mode is displayed, press the

MEAS key to return to the measurement screen.

4) Press the

MEAS key once to activate measure menu items. Press the

MEAS key multiple times to move right, and the ESC key multiple times

to move left until the

TEMP cell is highlighted.

5) Press the

UP, DOWN, LEFT, and RIGHT arrow keys to scroll the

highlighted value.

6)

Alternatively , press the ENTER key to display the Digits Edit Box.

7) Press the

UP and DOWN arrow keys to scroll the highlighted value.

8) Press the

LEFT and RIGHT arrow keys to scroll the digit locations.

9) Repeat steps

7 & 8 until the TEMP value is correctly displayed.

10) Press the

OK key to return to the measurement screen, or ESC to

cancel entering the

TEMP.

9.3 Semi Automatic Mode

The semi automatic mode gives the user an opportunity to initiate when the
temperature is read and automatically adjusted. This mode may be desirable to
those users want to use the automatic mode but want to retain some control as
to when the temperature is changed, rather than take for granted the temperature
is automatically changing. Some companies and application procedures may call
for manually initiating and physically viewing the change in temperature by