Performing a field calibration – Checkline TI-MINIMAX User Manual

Page 115

Advertising
background image

MINI-MAX Ultrasonic Bolt Tension Monitor

2) Use the

UP and DOWN arrow keys to scroll through the sub menu items

until

LOAD CAL MODE is highlighted.

3) Press the

LEFT and RIGHT arrow keys to scroll through the unit options

OFF, VECTOR (zero offset), REGRESSION, until the correct mode is

displayed.

4) Press the

MEAS key to retur n to the measurement mode.

Now that a group has been created and activated, and the load calibration mode
selected, we’re ready to perform the field calibration.

Performing a Field Calibration

1) Place the first sample bolt in the load apparatus.

Note: The following steps assume that the user has previously selected a

similar material type for the list of materials and the approximate length of

the bolt has been entered. Therefore, the MINI-MAX is currently setup and

ready to measure in elongation mode.

2) Apply a drop of couplant to the bolt or transducer, and attach it to one

end of the bolt. Rotate the transducer clockwise and counter clockwise

applying a small amount of pressure to eliminate any excess couplant

between the transducer and bolt. Be sure to always place the

transducer in the same location. This will help to eliminate any potential

measurement errors cause by changing the sound path.

3) Record the ultrasonic reference length at zero Load by pressing the

ENTER key.

Note: The reference lengths must be saved in column A in the group.