Adjusting the threshold using the hot menus – Checkline TI-MINIMAX User Manual

Page 134

Advertising
background image

Dakota Ultrasonics

128

user does not increase it at that time, when the fastener is put under a load, the
same peak jumps can occur. Even though the threshold can be adjusted at a
later time to correct the peak jump, errors can be overlooked during the
measurement process. Therefore, it’s always best to make these adjustments
when initially measuring the reference lengths.

The procedures to adjust the

Threshold are outlined below:

Adjusting the Threshold using the Hot Menus

1) Press the

MEAS key once to activate measure menu items. Press the

MEAS key multiple times to move right and the ESC key multiple times

to move left, until the

THR (threshold) cell is highlighted.

2) Press the

UP, DOWN, LEFT, and RIGHT arrow keys to scroll the

highlighted value.

3)

Alternatively, press the ENTER key to display the Digits Edit Box.

4) Press the

UP and DOWN arrow keys to scroll the highlighted value.

5) Press the

LEFT and RIGHT arrow keys to scroll the digit locations.

6) Repeat steps

4 & 5 until the THR value is correctly displayed.

7) Press the

OK key to return to the measurement screen, or ESC to

cancel entering the

THR.