Performing a probe zero calibration – Checkline TI-MINIMAX User Manual

Page 60

Advertising
background image

Dakota Ultrasonics

54

original. If the hardware has not changed, the user should not use this feature. If
the hardware has not changed, only the temperature value should be adjusted to
match the zero values.

Note: If the accessory temperature sensor is used, it

will automatically adjust the temperature value for the user, thus
eliminating the need for manual adjustment. Therefore, the corresponding
values should match.

The next two sections outline first, The procedure for calculating a probe zero
because the hardware has changed, and second, by adjusting the temperature
value because the hardware is the same and the values do not match.

Note: Only one of the next two sections should be selected and performed.

Performing a Probe Zero Calibration

Note: User has selected this section because the MINI-MAX, transducer,

or transducer cable has changed.

1) Press the

MENU key once to activate the menu items tab. Press the

MENU key multiple times to tab right, and the ESC key multiple times to

tab left, until the

CAL menu is highlighted and displaying the submenu

items.

2) Use the

UP and DOWN arrow keys to scroll through the sub menu

items until

ZERO OPTION is highlighted.

3) Press the

LEFT and RIGHT arrow keys to scroll the zero mode options

until

ONE POINT is displayed.

4) Use the

UP and DOWN arrow keys to scroll through the sub menu items

until

MEASURE ZERO is highlighted.